Generator:
- Type- Constant Potential HV, Input Line- 120 - 230
VAC
X-Ray Tube:
- Metal Ceramic, Water-cooled, Tungsten Target, End
grounded
- Rating= 120kV/10mA, 40kV/30mA, 1200 watts continuous
dissipation
- Focal Spot Size= 1mm x 1mm
Base Unit Assembly:
- Basic Space Requirements: 29” x 30”
- Vertical Travel: up to 68”
- Horizontal Arm Travel: 17” to 61” from center
- X-Ray Tube Movement: V&H 180 degrees
- Integrated Modular Components: Input power, HV generator,
Heat exchanger
Operator Control Console:
- Can be located up to 100’ (30 meters) from base
unit
- Service mode for system set-up and calibration -
key entry
Two Treatment Techniques:
- 100kV @ 8mA, 2.1mm Al. HVL or 70kV @ 10mA, 1.3mm
Al. HVL
- X-Ray output is 460 cGy @ 15cm SSD
Automatic Filter Change (Patent Pending):
- 2.1mm Al. HVL / 1.3 mm Al. HVL / Pb X-Ray Block
Automatic Warm-up Procedures:
- Automatically activated from the time of last exposure
- Four pre-programmed sequences
- Lead blocker automatically placed over x-ray tube
support
RAD Check:
- Direct radiation measurement of output and pre-treatment
verification
System Weight:
Accessories:
- The SRT 100 includes all the various sized Applicators
ranging from 1.5cm to 10.0cm, an Applicator Replacement
TIP Kit, a Service Spare Fuse/Hardware Kit, a Cable
Retractor Installation Kit and a set of Physicist
Calibration Fixtures.
Replaceable Safety Contact Shields:
- Applicator size specific. Visibility of treated
area.
Maximum Field Size:
Maximum Energy:
|